Sanyo's MCO-19M multi-gas incubator provides accurate control of oxygen (O2) and carbon-dioxide (CO2) levels combined with precisely regulated temperature.
By achieving a stable, uniform culturing environment, the incubator ensures ideal growth conditions and reproducibility of experimental results.
The MCO-19M also introduces the option of a hydrogen-peroxide decontamination system, which enables full decontamination and incubator cleaning to be completed in less than a morning.
Incorporating Sanyo's proprietary single-beam, dual-detector infrared CO2 sensor, the MCO-19M incubator delivers precise CO2 control and continuous zero calibration.
The solid-state CO2 sensor is not affected by humidity or temperature changes and has no moving parts, so reliability is improved.
O2 is measured using a solid-state zirconium sensor.
Sub-ambient O2 levels can be maintained from one to 18 per cent and enriched levels from 22 to 80 per cent, simultaneously with a CO2 range from zero to 20 per cent.
This gives the MCO-19M flexibility for use in a range of cell-culture applications.
A PID microprocessor provides accurate control of CO2 and O2 levels, ensuring rapid recovery without overshoot after frequent door openings, making it suitable for even the busiest laboratories.
The MCO-19M incubator utilises Sanyo's patented Direct Heat and Air Jacket conditioning system.
With multiple heating zones under microprocessor control, the temperature throughout the chamber is precisely regulated.
Users can therefore rely on getting the same results from every shelf.
Contamination prevention is another area where this new multi-gas incubator can be used.
The Incu Safe copper-enriched stainless-steel interior continuously inhibits growth of moulds, fungi, mycoplasma and bacteria.
With the optional patented Safecell UV system, airborne and water contaminants are eliminated without harming cell cultures.
The Sanyo Hydrogen Peroxide decontamination system option is fully verified, making it suitable for use in all laboratories, especially those operating to GMP/GLP.
With no need to remove or replace, then reinstate critical incubator components, this high-speed vaporised H2O2 process takes less than three hours including the initial cleaning.
Other features of the MCO-19M include an automatic gas-switchover system that changes from the primary to a secondary gas cylinder for either oxygen or nitrogen, and an optional secondary gas switchover system for CO2, ensuring continuity of supply.
There is also a newly designed multiple glass inner door system, which makes it unnecessary for the door to be fully opened when removing cell cultures.
This helps the conditions on other shelves remain more uniform and aids rapid recovery of the optimal culturing environment.
The MCO-19M multi-gas incubator is suited for life-science, clinical and pharmaceutical research, as well as highly regulated healthcare applications such as IVF and regenerative medicine.