The Rapidox 2100 can be used to accurately analyse oxygen and moisture levels during the welding of titanium.
There is little room for error when using shielding gas during titanium welding – 0.005% to be exact.
The American Welding Society (AWS) recommends measuring shielding gas purity to ensure weld quality.
Most titanium applications call for a 100 percent Argon shielding gas measured to be at least 99.995% pure with no more than 20 parts-per-million (ppm) of oxygen.
This is where the Rapidox 2100 oxygen gas analyser can be used to analyse the oxygen levels accurately.
The zirconia sensor on the Rapidox 2100 will measure oxygen down to 10-20ppm O2, easily covering the range recommended by the AWS.
With a fast response time of 4 seconds for a T90 measurement and a high accuracy of - 1% of range measured, the Rapidox 2100 can be used to ensure the shielding gas supply is fit for purpose.
The Rapidox 2100 is supplied with an external zirconia sensor connected to a 2m cable, which can be extended to 30m; this allows the user to take a measurement from distance if necessary.
The sensor can also withstand pressures of up to 10 bar and can be supplied with an auxiliary.
This allows the Rapidox to auto correct for any pressure changes in the shielding gas or to ensure the correct amount is displayed if the user is working with pressurised gas.
Alternatively some titanium welding applications may require dewpoint measurement.
Moisture can have a harmful effect on titanium welding as, similar to aluminium oxide, titanium oxide is hygroscopic so it absorbs moisture from the atmosphere.
Welding of titanium components with hydrated layers on the joint surfaces results in increased levels of gaseous hydrogen dissolved in the weld metal and subsequent pore formation upon solidification.
The Rapidox 2100 can also be fitted with an auxiliary sensor to measure H2O content from 20oC to -100oC.
This allows the user to ensure the level of moisture in the gas will not affect the weld.
The Rapidox 2100 also includes a Type K thermocouple input as standard for independent temperature measurements up to 1250°C.
The auxiliary input can be used to connect an optional pressure / vacuum / dewpoint sensor which can be displayed and data-logged simultaneously.
Standard features include:
- Fully programmable alarm circuits
- Programmable analogue outputs (0-5V & 4-20mA)
- Easy calibration (user selectable gases)
- RS232 / RS485 communications
- A full set of communications / data-logging software which is compatible with MS-Excel