The new 9 series Titrando range of autotitrators with Iconnect guarantees complete traceability of the analytical result to each component participating in the analysis.
Isolvotrode 6.0279.100 Iconnect and Itrode prevent the use of an incorrect electrode or an electrode with an expired calibration period or working life.
The 854 Iconnect is an electrode cable and measuring amplifier in one.
The built-in memory chip allows the storage of important sensor data such as article and serial numbers, calibration data, calibration history, working life and calibration validity period.
All the sensor data is read in automatically when the Itrode is connected to the Titrando.
This eliminates mix-ups or editing errors.
The electrode is identified automatically.
If the type of electrode is not the same as that defined in the method, then the user is informed.
This prevents the use of an incorrect electrode.
Monitoring functions allow the exclusion of electrodes with calibration data that lies outside the limits or with expired calibration periods.
If the sensor is used with different instruments or if the user wants to prevent inexperienced users from calibrating the electrode on their own instruments then the electrode can be calibrated on a different instrument under defined conditions.
The calibration data stored in the chip makes the electrode transferable; it does not need to be recalibrated each time it is used with a different instrument.
Despite its new digital measuring interface and intelligent sensors, the new range of Titrando autotitrators with combined analogue and digital measuring amplifier also supports conventional sensors.