Humidity is an increasingly important measurement and control parameter in a wide range of industrial and scientific processes as well as in the validation of product performance
At the very highest level, variations in ambient humidity can even affect the uncertainty of fundamental measurement standards; for example, the primary realisation of the kilogramme at Bureau International des Poids et Measures (BIPM), Paris, France, depends upon the accurate determination of atmospheric humidity using a cooled mirror dewpointmeter in order that the effect of air buoyancy can be corrected for.
Traceability of humidity measurement is currently limited to the range -75 to +82C dew point, or its equivalent in other engineering units.
This sounds small, but in concentration terms relates to the range 1 to 1,000,000 parts per million (dry basis).
Generally speaking the uncertainty of such standards is of the order of a few hundredths of a degree Celsius, or 0.5 to 1% of measured concentration. Uncertainty variances and sensor type selection.
The transfer of the measurement capability of a national standard across to a working industrial hygrometer can add significantly to that instrument's measurement uncertainty.
A polymer film relative humidity sensor used for measuring the dew-point temperature in a clean dry air line can, for example, have a measurement uncertainty of +/-100% of measured value, or approximately 7.5C frost point.
Such polymer film probes are commonly quoted as having 1 or 2% humidity uncertainties. Impedance type dew point sensors give a significantly better uncertainty as they are designed and calibrated in the unit of measurement.
A good impedance dew-point sensor will have an uncertainty of between 1 and 2C, which relates to an uncertainty in terms of concentration of about +/-15 %In order to reduce the uncertainty of measurement further, a fundamental cooled mirror measurement technique may be employed.
Cooled mirror dewpointmeters are widely used as transfer standard devices to provide the link between a national standard (or calibration laboratory) and a field measurement instrument.
A typical cooled mirror instrument will have a measurement uncertainty of 0.2 to 0.3C frost point at this level, which equates in volumetric humidity (concentration) terms to 1.5 to 2% of reading.
From the aforementioned range of uncertainties - from 100% of reading for a polymer film RH probe down to 2% of reading for a cooled mirror sensor at 250ppmv measured humidity - it can be seen that careful selection of the sensor type is just as important as the need to maintain traceability by regular calibration or validation.
The fact that an RH probe is 50 times less accurate than a dew point sensor at 250ppmv throws into question the decision to select this type of sensor for anything but trend measurement at low humidity levels.
Traceability and accreditation.
Traceability through calibration has been possible in the UK humidity industry for nearly 17 years.
The UK's national humidity standard, operated by the National Physical Laboratory (NPL), was commissioned in 1985 and continues to provide the fundamental reference, with direct traceability itself back to mass and temperature standards.
NPL's two-temperature humidity generator is validated periodically against a primary gravimetric hygrometer and is used to calibrate transfer standard dew-point hygrometers by reference against its own transfer standard hygrometer and reference temperature probes to measure critical loop temperature values.
Since 1986 Namas (now Ukas) has operated an accreditation service for humidity measurements, and the calibration laboratory of Michell Instruments was the first to be accredited in May of that year.
Continuous, unbroken accreditation has been maintained ever since and Michell continues to be the only UK manufacturer of cooled mirror and impedance dew-point hygrometers to have Ukas accreditation. Through the European Cooperation on Accreditation (EA), International Accreditation Forum (IAF), and International Laboratory Accreditation Co-operation (ILAC) this accreditation has global recognition.
With a wide range of impedance hygrometers and cooled mirror dewpointmeters, capable of operating over a range from -110 to +90C dew point, Michell can provide a measurement solution for almost any application, from critical corrections to fundamental mass balances to low cost measurement of compressed air lines.
In addition to Ukas accreditation, Michell's calibration laboratory offers commercial calibration of Michell and non-Michell products over the full humidity range, with traceability to NPL and Nist over the range -75 to +82C dew point.
Its expertise in humidity calibration also allows it to develop and install commercial humidity generators and calibration systems for major industrial users and standards institutes.