For measuring the surface area and porosity of a broad range of applications such as pharmaceuticals, ceramics, carbons, catalysts, paint and coatings, etc.
The Micromeritics Gemini V series surface area analysers retains the immensely successful, patented twin-tube design of the previous series of Gemini products.
This design assures fast and accurate dosing of the adsorptive gas without pressure overshoot and also negates free-space errors introduced by thermal gradient variations or by initial mismeasurement of free space.
This analytical capability, paired with embedded software, powerful new Windows controlled data archiving and reporting software, or Micromeritics's new Confirm 21 CFR Part 11 software, offers the user fast, superior results with a wide range of data presentation options.
Two models of Gemini are available.
The Gemini 2365 determines single-point and multipoint BET and Langmuir surface areas, total pore volume, and micropore analysis by the T-method.
The capability to determine statistical thickness surface area (STSA) of carbon blacks is also included. The Gemini 2380 has all the abilities of the Gemini 2365 and additionally can provide BJH pore volume distributions.
Up to one thousand isotherm data points may be collected.
The Gemini V series is ideal for measuring the surface area and porosity of a broad range of applications such as pharmaceuticals, ceramics, carbons, catalysts, paint and coatings, propellants, electronics, cosmetics, and aerospace materials.